Annual maintenance—provider, Introduction, Annual maintenance check list – AirSep Eclipse 5 20631679 User Manual
Page 38: Checking and replacing the power cartridge, Eclipse 5 monthly run-time procedure
Personal Ambulatory Oxygen System Provider Technical Manual
38
PN 20631679 Rev C
INTRODuCTION
Properly maintaining the Eclipse 5 will ensure longer life and higher performance. Minimum annual maintenance is required.
ANNuAL MAINTENANCE CHECK LIST
Perform the following maintenance procedures at least once a year or more often, as needed. The frequency of the periodic
maintenance should be based on the environment in which the Eclipse 5 is used.
• Replace air inlet filter
• Check Power Cartridge
• Inspect AC and DC Power Supply plugs and cords for damage. Replace as needed
• Verify that the AC and DC Power Supplies operate with the Eclipse 5
• Readjust the Pulse Dose Sensitivity setting between patients as needed
• Readjust the Rise Time setting between patients, as needed
• Replace 9V battery
• Check Universal Cart for functioning wheels and telescoping handle. Replace as needed
• Read and record hour meter
• Check flow rate, purity, Pulse Dose Mode and alarm functions
• Replace compressor intake filter
• Perform electrical safety test (Required only if used in a hospital or institutional setting; not required for home use)
• Replace HEPA filter
CHECKING AND REPLACING THE POWER CARTRIDGE
Typically, the Power Cartridge will require no routine maintenance beyond cleaning by the patient and calibration. SeQual
recommends that the Power Cartridge be replaced if there is physical damage to the cartridge enclosure or connector, or if the
battery’s recharge capacity, after calibration, is 20% less than published values at its given flow rate.
ECLIPSE 5 MONTHLy RuN-TIME PROCEDuRE
1. Power on Eclipse 5 using AC Power, DC Power, or Power Cartridge
2. Allow unit to run for a minimum of 2 hours
Annual Maintenance—Provider
CAUTION
The Eclipse 5 contains electrostatic sensitive components. Do not
open or handle except at a static free workstation. Do not remove
cover without ESD protection.
NOTE:
This procedure is performed to ensure and prolong the life of your Eclipse 5. This is also an ample
opportunity to calibrate the power cartridge of the Eclipse 5.