beautypg.com

AirSep Eclipse 5 20631679 User Manual

Page 47

background image

Personal Ambulatory Oxygen System Provider Technical Manual

47

PN 20631679 Rev C

To replace the compressor intake filter, follow these steps:

1. Cut the cable tie on the silicone tubes attached to the compressor intake filter.

Figure 24: Removing the Compressor Intake Filter.

2. Remove the silicone tubes attached to each end of the filter body.

3. Install a new filter by pushing each tube completely over the barb on the filter body. Verify

flow direction is correct.

4. Reconnect silicone tubes to the filter. Ensure that the feed tubes are not blocked, crimped

or kinked upon completion of the installation or the unit will alarm for low oxygen purity

after the warm-up cycle is complete.

Figure 25: Installing the Compressor Intake Filter.

5. Re-install two cable ties on silicone tubes.