10.3 Effect of cable temperature on the accuracy of probes HC2-C04 and

HC2-C05


Accuracy of the HC2-C04 and HC2-C05 probes is affected by the temperature of the cable separating the
sensors from the electronics.

Error on humidity:

Humidity measurement accuracy is specified at ±1.0 %RH when both the humidity sensor and the cable are at
23 °C. Depending both on the value of temperature and on the exposed length of cable, the probe exhibits a
predictable error (bias) as illustrated below:


Error on temperature:

Temperature measurement accuracy is specified at ±0.3 °C when both the temperature sensor and the cable
are at 23 °C. When the full length of the 2 meter probe cable is exposed to temperature values within the
range of -40 to 85 °C, the additional error is limited to 0.125 °C.

11 Accessories and parts

For accessories and parts such as the HW4 configuration software, extension cables, adapter cables,
calibration accessories and spare dust filters, please see document E-M-HC2-accessories

12 Supporting documents

Document File Name

Contents

E-M-HC2-accessories

Accessories and parts for probes, indicators and transmitters

Systematic error as a function of the length of cable

exposed to temperature