10.3 Effect of cable temperature on the accuracy of probes HC2-C04 and
HC2-C05
Accuracy of the HC2-C04 and HC2-C05 probes is affected by the temperature of the cable separating the
sensors from the electronics.
Error on humidity:
Humidity measurement accuracy is specified at ±1.0 %RH when both the humidity sensor and the cable are at
23 °C. Depending both on the value of temperature and on the exposed length of cable, the probe exhibits a
predictable error (bias) as illustrated below:
Error on temperature:
Temperature measurement accuracy is specified at ±0.3 °C when both the temperature sensor and the cable
are at 23 °C. When the full length of the 2 meter probe cable is exposed to temperature values within the
range of -40 to 85 °C, the additional error is limited to 0.125 °C.
11 Accessories and parts
For accessories and parts such as the HW4 configuration software, extension cables, adapter cables,
calibration accessories and spare dust filters, please see document E-M-HC2-accessories
12 Supporting documents
Document File Name
Contents
E-M-HC2-accessories
Accessories and parts for probes, indicators and transmitters
Systematic error as a function of the length of cable
exposed to temperature
-10
-8
-6
-4
-2
0
2
4
6
8
10
0
20
40
60
80
100
120
140
160
180
200
Cable Length [cm]
E
rro
r [%
R
H
]
-40 °C
-20 °C
0 °C
23 °C
40 °C
60 °C
85 °C