MEASUREMENT LOOP VALIDATION

AirChip 3000 Functions

Description

► Simulator mode

Used to make the probe generate fixed values for the
humidity, temperature and calculated parameter. Can be
configured, enabled or disabled

SAFEGUARDS

AirChip 3000 Functions

Description

► Device write protection

Used to protect the probe with a password to prevent
unauthorized digital access by a digital user. Can be
configured, enabled or disabled

PROCESS PROTECTION / PROTECTION OF OTHER DEVICES

AirChip 3000 Functions

Description

► Limit humidity output to 100 %RH

Used to prevent the humidity signal from exceeding
100 %RH when condensation forms on the sensor. Can be
enabled or disabled

► Out-of-limit value alarm

Used to specify the normal range for humidity, temperature
and the calculated parameter depending on the user
application. Can be configured, enabled or disabled

Out-of-limit values trigger a digital alarm

► Bad sensor alarm

Built-in function. Cannot be disabled

A bad humidity or temperature sensor triggers a digital alarm

► Fail safe mode

Used to specify a "safe" fixed value for humidity and for
temperature in the event of a sensor failure. Can be
configured, enabled or disabled

4.2 Factory default settings


Note: Configuration of the probe by the user and access to its functions requires a PC with the ROTRONIC
HW4 software (version 2.1.1 or higher) installed. For connecting the HC2 probe to a PC, use any of the
following digital adapter cables AC3001, AC3002 or AC3005 (see document E-M-HC2-accessories).

Configurable Settings

Factory default

Unit system (Metric or English)

Metric

Psychrometric calculation

None

Output 1 parameter, scale and unit

Humidity: 0…100%RH

Output 2 parameter, scale and unit

Temperature: -40…60 ˚C

Communication protocol

RO-ASCII

RS-485 address

0

Device name

Probe type