HygroClip 2 (HC2) Humidity Temperature
Probes: User Guide

Instruction Manual

Document Type

Page 26 of 31

Document title

© 2009-2013; Rotronic AG E-M-HC2 Probes-V1_26

10 Technical data

10.1 Specifications

General

Device type

Humidity temperature probe

Mechanical configuration

See models

Power supply and connections

Supply voltage (VDD)

3.3 … 5 VDC ± 0%

Recommended supply voltage

3.3 VDC

Supply voltage during factory
tests and adjustment

3.3 VDC

Supply voltage stability / ripple

See System Accuracy, effect of VDD

Nominal current consumption

< 4.5 mA at VDD = 3.3…5 VDC

Maximum start-up current unlimited

< 50mA during 2µs

Minimum start-up current

8 mA during 2ms and 5 mA during 2s

Maximum current spike during
operation

1 mA during maximum 2µs

Polarity protection

Mechanical only (keyed connector)

Humidity measurement

Sensor ROTRONIC

Hygromer

®

IN1

Measuring range

0…100 %RH

Measurement accuracy at 23 °C
with 3.3 VDC supply voltage

±0.8 %RH (w. standard adjustment profile)
±1.0 %RH for models HC2-C04 and HC2-C05 (see also §10.3)

Repeatability 0.3

%RH

Long term stability

< 1 %RH / year

Sensor time constant

Typical 10 sec, 63% of a 35 to 80 %RH step change (1m/sec air flow at sensor)


Temperature measurement

Sensor

Pt100 RTD, IEC 751 class A

Measuring range

-100…200 °C (see also environmental limits)

Measurement accuracy at 23 °C
with3.3 VDC supply voltage

±0.1 °C
±0.3 °C for models HC2-C04 and HC2-C05 (see also §10.3)

Repeatability 0.05°C
Long term stability

< 0.1°C / year

Sensor time constant

Typical 4 sec, 63% of a step change (1m/sec air flow at sensor)

Calculated parameters

Psychrometric calculations

Dew or frost point (user configurable)

Start-up time and data refresh rate

Start-up time

1.5s (typical)

Data refresh rate

1.0s (typical) – when not calculating any parameter