Conformity to standards

CE / EMC immunity

EMC Directive 2004/108/EG: EN 61000-6-1: 2001, EN 61000-6-2: 2005
EN 61000-6-3: 2005, EN 61000-6-4: 2001 + A11

Solder type

Lead free (RoHS directive)

FDA / GAMP directives

compatible

Environmental limits

Storage and transit

-50…+100 °C / 0…100 %RH, non condensing

Operating limits at electronics

-40 … +100 °C
0…100 %RH, non condensing

Temperature limits at sensor

See Models

Maximum humidity at sensor

100 %RH up to 80 °C (176 °F)
75 %RH at 100 °C (212 °F)
45 %RH at 125 °C (260 °F)
15 %RH at 150 °C (302 °F)

Maximum air velocity at probe

20 m/s (3,935 ft /min)

Critical environments

Humidity sensor: as per DV04-14.0803.02 - Critical chemicals

10.2 Dew point accuracy

The HC2 probes can be configured to calculate either the dew point or frost point based on the measurement
of relative humidity and temperature. The accuracy of this conversion varies, depending on the humidity and
temperature conditions as shown in the graph below:


Example: at a temperature of 20 ˚C, a dew point value of -37 ˚C is measured with an accuracy of
± 1.0 ˚C or better.