beautypg.com

E-m-hc2 probes-v1_26, 4user configurable settings and functions, 1 function overview – ROTRONIC HC2 User Manual

Page 15

background image

E-M-HC2 Probes-V1_26

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroClip 2 (HC2) Humidity Temperature
Probes: User Guide

Instruction Manual

Document Type

Page 15 of 31

Document title

© 2009-2013; Rotronic AG E-M-HC2 Probes-V1_26

4

User configurable settings and functions


The HC2 probe ships configured as per the factory defaults:

o

Users who intend to use the probe as a stand-alone analog device can use the probe as any conventional

humidity and temperature probe. A connecting cable with active electronics may be required to power the
probe. Most such users will never have to use the probe configurable settings and functions

o

Users who purchased a probe for use with another ROTRONIC instrument can also use the probe right

away and do not necessarily have to use the probe configurable settings and functions

o

Users that intend to use the probe as a stand-alone digital device typically need to use a connecting cable

with active electronics and generally need to configure either the probe or the PC


Making use of the probe configurable settings and functions is entirely up to the user and the appropriate settings
depend on the user application. We have provided below a short description of the probe functions and also
indicated the factory default settings.

4.1 Function overview

MEASUREMENT ACCURACY AND RELIABILITY

AirChip 3000 Functions

Description

► Humidity / temperature adjustment

o

1-point or multi-point humidity calibration or adjustment

o

1-point or 2-point temperature calibration or adjustment

o

Generate a time stamp for calibrations and adjustments

o

Retain and view last adjustment date and adjustment values

o

Generate calibration and adjustment protocols

► Automatic humidity sensor test
and optional drift compensation

Tests the humidity sensor for drift caused by contaminants and
can be used to automatically apply a correction. The test is
automatically carried out at regular intervals of time. Can be
configured, enabled, or disabled

The humidity sensor status can be verified either with the HW4
software and is shown as Good, SQ-tuned (corrected for drift) or
Bad (defective)

► Data recording

The data recording function differs from a true data logging
function in the sense that the AirChip 3000 does not time stamp
the data. This data recording function can be used to investigate
events such as a sensor malfunction as well as to retrieve data
that would otherwise be lost

o

Start or stop data recording - up to 2000 value pairs (%RH

and temperature). Starting a recording session erases all
previously recorded data

o

The recording mode and log interval can be specified

o

When the probe is powered off, the recording session is

paused but not ended As long as the recording session has
not been ended, the probe automatically resumes recording
data when powered up again

o

The recorded data can be downloaded to a PC with the

HW4 software, time stamped and viewed