E-m-hc2 probes-v1_26, 6 connector, 7 sensor protection (dust filter) – ROTRONIC HC2 User Manual
Page 14

E-M-HC2 Probes-V1_26
Rotronic AG
Bassersdorf, Switzerland
Document code
Unit
HygroClip 2 (HC2) Humidity Temperature
Probes: User Guide
Instruction Manual
Document Type
Page 14 of 31
Document title
© 2009-2013; Rotronic AG E-M-HC2 Probes-V1_26
3.6 Connector
All standard probe models of the HC2 series use the same keyed connector (male). The connector is located
either directly on the probe body or at the end of a cable. The probe connector is secured to the matching female
connector with a threaded collar.
Pin-out diagram
3.7 Sensor protection (dust filter)
Most probes of the HC2 series are supplied with a filter to protect the sensors against dust particles and high
air velocity. Depending on the probe model the following types of filter materials are used:
Material
Maximum temperature
Notes
Polyethylene
100 °C (212 °F)
Recommended filter material for all applications
below 100 °C. Good response and good
protection against fine dust particles. No water
absorption or retention
Teflon
200 °C (392 °F)
Good protection against fine dust particles and
salt (marine environments). Moderately slow
response
Stainless steel wire
mesh
200 °C (392 °F)
Provides fastest response time. Not
recommended in environments with fine dust
particles (clogging) and in bioactive environments
Stainless steel sintered
filter
200 °C (392 °F)
Good response at low humidity levels. Do not use
at high humidity levels. Provides best protection
against abrasive particles
Note: Depending on the probe model, the dust filter is either automatically included with the probe or it must
be ordered separately (see Models).
Probe connector (7-pin male – looking at probe)
Pin #
Wire Color
Name
Function
1
Green
VDD (+)
3.3 to 5 VDC
2
Grey
GND
Power and digital signal
3 Red
RXD
UART
4 Blue
TXD
UART
5
White
Out 1 analog (+)
Humidity 0…100%RH (default)
6
Brown
Out 2 analog (+)
Temperature -40…60°C (default)
7
Yellow
AGND
Analog signal ground