7.2.1

Minimum load requirements for the probe voltage outputs


The following requirements apply to any external device or circuit connected to the probe voltage outputs:



















In the situation where the external device uses an internal pull-up resistor the value of this resistor should
meet the requirements shown below. It is also necessary to add a pull-down resistor RL connected to ground
in order to be able to read 100% of the range of the probe voltage output.





















HFC2 output signal

Input resistance RL

0…1V >=1kOhm

HC2 signal

VCC

R pull-up

RL

0…1V

3.3V

≥ 250 kOhm

1 kOhm

HC2 signal

VCC

R pull-up

RL

0…1V

5.0V

≥ 400 kOhm

1 kOhm

HC2 signal

VCC

R pull-up

RL

0…1V

10.0V

≥ 1 MOhm

1 kOhm

RL

GND

HC2

Voltage Output

R Pull-up

RL

GND

VCC

HC2

Voltage Output