7.2.1
Minimum load requirements for the probe voltage outputs
The following requirements apply to any external device or circuit connected to the probe voltage outputs:
In the situation where the external device uses an internal pull-up resistor the value of this resistor should
meet the requirements shown below. It is also necessary to add a pull-down resistor RL connected to ground
in order to be able to read 100% of the range of the probe voltage output.
HFC2 output signal
Input resistance RL
0…1V >=1kOhm
HC2 signal
VCC
R pull-up
RL
0…1V
3.3V
≥ 250 kOhm
1 kOhm
HC2 signal
VCC
R pull-up
RL
0…1V
5.0V
≥ 400 kOhm
1 kOhm
HC2 signal
VCC
R pull-up
RL
0…1V
10.0V
≥ 1 MOhm
1 kOhm
RL
GND
HC2
Voltage Output
R Pull-up
RL
GND
VCC
HC2
Voltage Output