Command
Name
Flow Meter
Linearizer

LCD Back
Light**

Noise
Reduction
Filter

Read
EEPROM
Memory
Write
EERPROM
Memory

Description
Enable/disable flow meter
linearizer

Display/change LCD back light
settings
Note: LCD back light settings has
to be in the range: )% ≤ BL ≤ 80%

Noise Reduction Filter parameter
settings. Following arguments
supported:

T - Running Average Time

Interval (0-99 seconds)

N -Running Average Number

of samples (1-32)

Reads the value in the specified
memory location.

Writes the specified value to the
specified memory location. Use
Carefully, can cause unit to
malfunction.
(Note: Some addresses are
write protected!)

No.
17

18

19

20

21

Command
FL

BL

NR

MR

MW

Argument 1

E or D
No Argument
(Returns current
linearizer settings)
(New LCD back
light value) 0 to 80%
No Argument
(Returns current
LCD back light
settings)
T (time interval)
N (number of
samples)
S (status)

0 to 100 (memory
index table)

20 to 100 (memory
index table)

Argument 2

Value

Argument 3

Argument 4

Response
FL:
Example: FL:E
FL:
Example: FL:E

BL:
Example: BL:50
BL:
Example: BL:50

NRT:5
NRN:5

NR:5,5

MW,XXX,
where:
XXX = table index
Example:
MW,100,”Meter#6”

UART Error Codes:
1 - Not Supported Command or Back Door is not enabled.
2 - Wrong # of Arguments.
3 - Address is Out of Range (MR or MW commands).
4 - Wrong # of the characters in the Argument.
5 - Attempt to Alter Write Protected Area in the EEPROM.
6 - Proper Command or Argument is not found.
7 - Wrong value of the Argument.
8 - Reserved.
9 - Manufacture specific info EE KEY (wrong key or key is disabled).

Diagnostic events codes and bit position:
1. CPU Temp. High

0

2. Flow > 125% FS

1

3. High Flow Alarm

2

4. Low Flow Alarm

3

5. High Temperature Alarm

4

6. Low Temperature Alarm

5

7. Temperature Above Limit

6

8. Temperature Below Limit

7

9. Main Totalizer ≥ Limit

8

10. Pilot Totalizer ≥ Limit

9

11. EEPROM Failure

A

12. DC/DC converter Voltage too High

B

13. DC/DC converter Voltage too Low

C

14. Communication Error

D

15. Reserved

E

16. Fatal ERROR

F