Badger Meter M-Series M4000 User Manual
Page 22
22
Press +, - or E once.
7. Main Screen appears. Press E again.
8. PASSWORD??? 01 screen reactivates. Input the correct
password.
Set Flow Rate - Unit of Measure and Totalizer Unit of
Measure
1. With Main Screen up, press E.
2. With MAIN MENU 00 active, press + twice
(Measurements) and then E.
3. On the MEASUREMENTS 20 screen press + once to
place cursor at Flow Unit. Press E.
4. From FLOW UNITS 21 press + or - to select a fl ow unit.
Press E.
5. This type of screen shows for two seconds.
6. The screen returns to MEASUREMENTS 20. With the
cursor at Exit this Menu, press + twice to move cursor to
Totalizer Unit. Press E.
7. From TotalizerUnit 22 press + or - to select a totalizer unit.
Press E.
8. This type of screen shows for two seconds.
9. Screen goes back to MEASUREMENTS 20.
Set Full Scale Flow Rate Value
1. Press + three times to move the cursor to FullScaleFlow.
Press E.
2. At FullScaleFlow 23 input the proper fl ow. Select E to
move cursor as described earlier.
3. Screen returns to MEASUREMENTS 20.
SET LOW FLOW CUTOFF
1. Press + four times, to place cursor at LowFlowCutOff.
Press E.