Input signals – Alicat M Series Mass Flow Meter User Manual
Page 9
9
INPUT SIGNALS
Analog Input Signal
Apply analog input to Pin 4 as shown on page 8.
For 6 Pin Locking Connector, DB9 and DB15 Pin-outs, see pages 58 to 70.
For PROFIBUS Pin-out, see page 56.
Standard 0-5 Vdc is the standard analog input signal. Apply the 0-5 Vdc input
signal to pin 4, with common ground on pin 8.
Opti onal 0-10 Vdc: If specifi ed at ti me of order, a 0-10 Vdc input signal can be
applied to pin 4, with common ground on pin 8.
Opti onal 4-20 mA: If specifi ed at ti me of order, a 4-20 mA input signal can be
applied to pin 4, with common ground on pin 8.
NOTE: This is a current sinking device. The receiving circuit is essenti ally a
250 ohm resistor to ground.
NOTE: 4-20 mA output requires at least 15 Vdc power input.
Electrical Connecti ons and Basic Wiring
htt p://www.alicat.com/support/instructi onal-videos/
A remote tare can be achieved by momentarily grounding pin 4 to tare
as shown above.
CAUTION! D� ��� ����e�� ��is �e�i�e �� “���� ���ere�’”
s�s�e�s, �s ��is �i�� �es�r�� ��r�i��s �� ��e �ir��i�r� ��� ��i�
��e ��rr����. I� ��� ��s� i��er���e �i�� e�is�i�� ���� ���ere�
s�s�e�s, �����s �se � si���� is�����r ��� � se��r��e ���er s�����.
4
8