beautypg.com

Alicat M Series Mass Flow Meter User Manual

Page 24

background image

24

Tareing via RS-232 / RS-485:
Tare –Tareing (or zeroing) the flow meter provides it with a reference point for

zero flow. This is a simple but important step in obtaining accurate measurements.

It is good practice to “zero” the flow meter each time it is powered up. A unit may

be Tared by following the instructions on page 10 or it may be Tared via RS-232

/ RS-485 input.
To send a Tare command via RS-232 / RS-485, enter the following strings:
In Polling Mode: Address$$V (e.g. B$$V)
Changing From Streaming to Polling Mode:
When the meter is in the Streaming Mode (RS-485 units do not have a streaming

mode), the screen is updated approximately 10-60 times per second (depending

on the amount of data on each line) so that the user sees the data essentially in

real time. It is sometimes desirable, and necessary when using more than one

unit on a single RS-232 line, to be able to poll the unit.
In Polling Mode the unit measures the flow normally, but only sends a line of data

when it is “polled”. Each unit can be given its own unique identifier or address.

Unless otherwise specified each unit is shipped with a default address of capital

A. Other valid addresses are B thru Z.
Once you have established communication with the unit and have a stream of

information filling your screen:
1. Type *@=A followed by “Enter” (or using the RS-232 / RS-485 communication

select menu, select A as identifier and exit the screen) to stop the streaming

mode of information. Note that the flow of information will not stop while

you are typing and you will not be able to read what you have typed. Also,

the unit does not accept a backspace or delete in the line so it must be typed

correctly. If in doubt, simply hit enter and start again. If the unit does not

get exactly what it is expecting, it will ignore it. If the line has been typed

correctly, the data will stop.

2. You may now poll the unit by typing A followed by “Enter”. This does an

instantaneous poll of unit A and returns the values once. You may type A

“Enter” as many times as you like. Alternately you could resume streaming

mode by typing *@=@ followed by “Enter”. Repeat step 1 to remove the

unit from the streaming mode.

3. To assign the unit a new address, type *@=New Address, e.g. *@=B. Care

should be taken not to assign an address to a unit if more than one unit is on

the RS-232 / RS-485 line as all of the addresses will be reassigned. Instead,

each should be individually attached to the RS-232 / RS-485 line, given an

address, and taken off. After each unit has been given a unique address, they

can all be put back on the same line and polled individually.