We recommend the use of in-line sintered filters to prevent large

particulates from entering the measurement head of the instrument.

Suggested maximum particulate sizes are as follows:

5 microns for units with FS flow ranges of 0-1 sccm or less.

20 microns for units with FS flow ranges between 0-2 sccm and 0-1 slpm.

50 microns for units with FS flow ranges of 0-1 slpm or more.

Connecting Fittings and Filters

http://www.alicat.com/support/instructional-videos/

PRESSURE

Maximum operating line pressure for M-Series units is 145 psig (1 MPa).
If the line pressure is higher than 145 psig (1 MPa), use a pressure regulator

upstream from the flow meter to reduce the pressure to 145 psig (1 MPa) or less.
Maximum operating line pressure for WHISPER units is 50 psig.

Exceeding the maximum specified line pressure may cause permanent

damage to the solid-state differential pressure sensor.

Do Not subject an M-Series Differential Pressure sensor to

upstream-downstream pressure differentials exceeding 75 PSID.
Do Not subject a WHISPER
Differential Pressure sensor to

upstream-downstream pressure differentials exceeding 15 PSID.

While high static pressure will typically not damage the dp sensor, sudden

pressure “spikes” can result in complete failure of the sensor.

A common cause of this problem is instantaneous application of

high-pressure gas as from a snap acting solenoid valve either upstream

or downstream of the meter. If you suspect that your pressure sensor is

damaged please discontinue use of the meter and contact Alicat.