Db15 pin-outs, Db15 – pin-out alicat style – Alicat M Series Mass Flow Meter User Manual
Page 65
65
DB15 – Pin-Out Alicat Style
Pin Number
Function
1
Ground
2
Primary Analog Signal Output
3
Ground
4
N/C
5
Power Supply (+Vdc)
6
N/C
7
N/C
8
Analog Tare (meters — when grounded)
Analog Set-Point Input (controllers)
9
Power Supply Common
10
Ground
11
Secondary Analog Signal Output / fi xed 5.12Vdc
12
N/C
13
RS-232 RX (receive) or RS-485 –
14
Ground
15
RS-232 TX (send) or RS-485 +
Check your device’s calibration certifi cate and user manual for the actual electrical input/
output requirements, as all instruments are custom confi gured to some extent.
NOTE: Pins 1, 3, 9, 10 and 14 are connected together inside of the device and are
common grounding points.
N/C = Not Connected/Open (can be used for custom pin assignments – please consult factory).
The following pin-out chart describes the safest and generally compatible arrangement
when connecting a non-Alicat DB15 wire to a
DB15 equipped Alicat. Not all features may be
available between brands, but the common denominators are featured in our DB15 offerings,
along with some options for customization.
DB15 Pin-Outs
If your instrument was ordered with a DB15 connection,
be sure to check the Calibration Label on the device and
reference the appropriate pin-out diagram.