Main – Alicat M Series Mass Flow Meter User Manual
Page 15
15
MAIN
This mode defaults on power up, with
mass flow as the primary displayed
parameter.
The following parameters are displayed in
the Main mode.
Gas Absolute Pressure: This sensor
references hard vacuum and reads
incoming pressure both above and below
local atmospheric pressure. This parameter
is moved to the primary display by pushing
the button above PSIA.
The engineering unit associated with
absolute pressure is pounds per square
inch absolute (psia). This can be converted to gage pressure (psig) by subtracting
local atmospheric pressure from the absolute pressure reading:
PSIG = PSIA – (Local Atmospheric Pressure)
Gas Temperature: M-Series flow meters
measure the incoming temperature of the
gas flow. The temperature is displayed
in degrees Celsius (°C). This parameter is
moved to the primary display by pushing
the button above °C.
Pushing the button again allows you to
select 0C (Celsius), 0K (Kelvin), 0F (Fahrenheit)
or 0R (Rankine) for the temperature scale.
To select a temperature scale, use the LN-UP
and LN-DN buttons to position the arrow in
front of the desired scale.
Press SET to record your selection and return
to the MAIN display. The selected temperature scale will be displayed on the screen.
Tare: Pushing the TARE V button tares the flow meter and provides it with a
reference point for zero flow. This is an important step in obtaining accurate
measurements. It is best to zero the flow meter each time it is powered up. If
the flow reading varies significantly from zero after an initial tare, give the unit a
minute or so to warm up and re-zero it.
If possible, zero the unit near the expected operating pressure by positively blocking
the flow downstream of the flow meter prior to pushing the TARE button.
Zeroing the unit while there is any flow will directly affect the accuracy
by providing a false zero point. If in doubt about whether a zero flow
condition exists, remove the unit from the line and positively block both ports
before pressing the TARE button. If the unit reads a significant negative value