beautypg.com

Alicat M Series Mass Flow Meter User Manual

Page 47

background image

47

Technical Data for WHISPER Low Pressure Drop Mass Flow Meters

0 to 0.5 sccm Full Scale through 0 to 500 slpm Full Scale

Standard Operating Specifications (Contact Alicat for available options.)

Performance

Whisper Mass Flow Meter

Accuracy at calibration conditions

after tare

± (0.8% of Reading + 0.2% of Full Scale)

High Accuracy at calibration conditions

after tare

± (0.4% of Reading + 0.2% of Full Scale)

High Accuracy option not available for units ranged under 5 sccm or over 500

slpm.

Accuracy for Bidirectional Meters

at calibration conditions after tare

± (0.8% of reading + 0.2% of total span from positive full scale to negative full scale)

Repeatability

± 0.2% Full Scale

Zero Shift and Span Shift

0.02% Full Scale / ºCelsius / Atm

Operating Range / Turndown Ratio

0.5% to 100% Full Scale / 200:1 Turndown

Maximum Measurable Flow Rate

128% Full Scale

Typical Response Time

10 ms (Adjustable)

Warm-up Time

< 1 Second

Communications / Power

Whisper Mass Flow Meter

Monochrome LCD or Color TFT

Display with integrated touchpad

Simultaneously displays Mass Flow, Volumetric Flow, Pressure and Temperature

Digital Output Signal

1

Options

RS-232 Serial / RS-485 Serial / PROFIBUS

3

Analog Output Signal

2

Options

0-5 Vdc / 1-5 Vdc / 0-10 Vdc / 4-20 mA

Optional Secondary Analog Output

Signal

2

0-5 Vdc / 1-5 Vdc / 0-10 Vdc / 4-20 mA

Electrical Connection Options

8 Pin Mini-DIN / 9-pin D-sub (DB9) / 15-pin D-sub (DB15) / 6 pin locking

Supply Voltage

7 to 30 Vdc (15-30 Vdc for 4-20 mA outputs)

Supply Current

0.040 Amp (+ output current on 4-20 mA)

1. The

Digital Output Signal communicates Mass Flow, Volumetric Flow, Pressure and Temperature

2. The

Analog Output Signal and Optional Secondary Analog Output Signal communicate your choice of Mass

Flow, Volumetric Flow, Pressure or Temperature

3. If selecting

PROFIBUS, no analog signal is available. PROFIBUS units do not have the display. See PROFIBUS

specifications for PROFIBUS supply voltages and currents (page 48) .

Operating Conditions

Whisper Mass Flow Meter

Mass Reference Conditions (STP)

25ºC & 14.696 psia (standard — others available on request)

Operating Temperature

−10 to +50 ºCelsius

Humidity Range (Non–Condensing)

0 to 100%

Maximum Pressure

50 psig

1

Higher line pressures available, please contact Alicat.

Mounting Attitude Sensitivity

None

Ingress Protection

IP40

Wetted Materials

303 & 302 Stainless Steel, Viton®, Silicone RTV (Rubber), Glass Reinforced

Nylon, Aluminum

If your application demands a different material, please contact Alicat.

1. Do Not subject a WHISPER Differential Pressure sensor to upstream-downstream pressure differentials exceeding 15 PSID.

Full Scale Flow

Mass Meter

Pressure Drop at FS Flow

(psid)venting to atmosphere

Mechanical Dimensions

Process Connections

1

0.5 sccm to 2 sccm

0.06

3.9”H x 2.4”W x 1.1”D

M-5 (10-32) Female Thread

2

5 sccm to 20 sccm

0.07

50 sccm

0.07

4.1”H x 2.4”W x 1.1”D

1/8” NPT Female

100 sccm to 200 sccm

0.06

500 sccm

0.07

1 slpm to 5 slpm

0.07

10 slpm

0.08

4.3”H x 2.7”W x 1.1”D

1/4” NPT Female

20 slpm

0.25

4.4”H x 4.0”W x 1.6”D

40 slpm

0.12

5.0”H x 4.0”W x 1.6”D

1/2” NPT Female

50 slpm

0.14

5.0”H x 4.0”W x 1.6”D

3/4” NPT Female

100 slpm

0.24

250 slpm

0.60

500 slpm

0.39

5.3”H x 5.2”W x 2.9”D

3/4” NPT Female

1. Compatible with Beswick®, Swagelok® tube, Parker®, face seal, push connect and compression adapter fittings. VCR and

SAE connections upon request.

2. Shipped with M-5 (10-32) Male Buna-N O-ring face seal to 1/8” Female NPT fittings.

Range Specific Specifications