Option: Totalizing Mode

Meters can be purchased with the Totalizing Mode option. This option adds an

additional mode screen that displays the total flow (normally in the units of the

main flow screen) that has passed through the device since the last time the

totalizer was cleared.

The Totalizing Mode screen is accessed by

pushing the TOTAL button on the MAIN display.
TOTAL/TIMER: Pushing the TOTAL/TIMER

button will cycle the large numbers on the

display between total mass and time elapsed.
Rollover – The customer can also specify at the

time of order what the totalizer is to do when

the maximum count is reached. The following

options may be specified:
No Rollover – When the counter reaches the

maximum count it stops counting until the

counter is cleared.

Rollover – When the counter reaches the maximum count it automatically rolls

over to zero and continues counting until the counter is cleared.
Rollover with Notification – When the counter reaches the maximum count

it automatically rolls over to zero, displays an overflow error, and continues

counting until the counter is cleared.

TOTAL MASS: The counter can have as many as seven digits. At the time of order,

the customer must specify the range. This directly affects the maximum count. For

instance, if a range of 1/100ths of a liter is specified on a meter which is totalizing in

liters, the maximum count would be 99999.99 liters. If the same unit were specified

with a 1 liter range, the maximum count would be 9999999 liters.
ELAPSED TIME: The small numbers below the mass total show the elapsed time

since the last reset in hours, minutes and seconds. The maximum measurable

elapsed time is 9999 hours 59 minutes 59 seconds. The hours count resets when

RESET is pushed, an RS-232 or RS-485 clear is executed or on loss of power. Press

ELAPSED TIME to show this as the primary display.
RESET – The counter can be reset to zero at any time by pushing the RESET

button. To clear the counter via RS-232 or RS-485, establish serial communication

with the meter or controller as described in the RS-232 or RS-485 section of the

manual. To reset the counter, enter the following commands:

In Polling (addressable) Mode: Address$$T (e.g. B$$T )

MENU

SCCM

+0.0

SCCM

TOTAL/

TIMER

BACK

RESET

0000:00:00

00000.00