Maintenance and Recalibration

General: M-Series Flow Meters require minimal maintenance. They have no

moving parts. The single most important thing that affects the life and accuracy

of these devices is the quality of the gas being measured. The meter is designed

to measure CLEAN, DRY, NON-CORROSIVE gases.
Moisture, oil and other contaminants can affect the laminar flow elements.

We recommend the use of in-line sintered filters to prevent large particulates

from entering the measurement head of the instrument. Suggested maximum

particulate sizes are as follows:

5 microns for units with FS flow ranges of 0-1 sccm or less.

20 microns for units with FS flow ranges between 0-2 sccm and 0-1 slpm.

50 microns for units with FS flow ranges of 0-1 slpm or more.

Recalibration: The recommended period for recalibration is once every year.

A label located on the back of the meter lists the most recent calibration date.

The meter should be returned to the factory for recalibration within one year

from the listed date. Before calling to schedule a recalibration, please note the

serial number on the back of the meter. The Serial Number, Model Number, and

Date of Manufacture are also available on the Model Info display (page 21).
Cleaning: M-Series Flow Meters require no periodic cleaning. If necessary, the

outside of the meter can be cleaned with a soft dry cloth. Avoid excess moisture

or solvents.