POWER AND SIGNAL CONNECTIONS

Power can be supplied to your meter through either the power jack (power jack not

available on CSA/ATEX approved devices) or the 8 pin Mini-DIN connector.
An AC to DC adapter which converts line AC power to DC voltage and current as

specifi ed below is required to use the power jack.
Meters require a 7-30 Vdc power supply with a 2.1 mm female positi ve center

plug capable of supplying at least 100mA.

Note: 4-20mA analog output requires at least 15 Vdc.

Standard 8 Pin Mini-DIN Pin-Out

Pin Functi on

Mini-DIN

cable color

1

Not Connected

(or opti onal 4-20mA Primary Output Signal)

Black

2

Stati c 5.12 Vdc [or opti onal Secondary Analog Output (4-20mA,

5Vdc, 10Vdc) or Basic Alarm]

Brown

3

Serial RS-232RX / RS-485(–) Input Signal (receive)

Red

4

Meters/Gauges = Remote Tare (Ground to Tare)

Controllers = Analog Set-Point Input

Orange

5

Serial RS-232TX / RS-485(+) Output Signal (send)

Yellow

6

0-5 Vdc (or opti onal 0-10 Vdc) Output Signal

Green

7

Power In (as described above)

Blue

8

Ground (common for power, digital communicati ons, analog signals

and alarms)

Purple

Note: The above pin-out is applicable to all the fl ow meters and controllers with the

Mini-DIN connector. The availability of diff erent output signals depends on the opti ons

ordered. Opti onal confi gurati ons are noted on the unit’s calibrati on sheet.

CAUTION! D� ��� ����e�� ���er �� �i�s 1 ��r���� 6 �s �er���e��

�����e ��� ����r!

It is common to mistake Pin 2 (labeled 5.12 Vdc Output) as the standard 0-5 Vdc

analog output signal. In fact Pin 2 is normally a constant 5.12 Vdc that refl ects

the system bus voltage and can be used as a source for the set-point signal.
For 6 Pin Locking Industrial Connector, DB9, and DB15 Pin-outs, see pages

58 to 70. For PROFIBUS Pin-out, see page 56

1

2

3

4

5

6

7

8