9

MOUNTING

All L-Series Meters and LC-Series Controllers have mounting holes for

convenient mounting to flat panels. The sizes and dimensions for the mounting

holes are shown on pages 51-55.
Meters are position insensitive and can be mounted in any orientation.
Small valve controllers (LC-Series) can usually be mounted in any position.
Large valve controllers (LCR-Series) should be mounted so that the valve

cylinder is vertical. Mounting a large valve controller in another position

increases the risk of leakage when the controller is being held closed by the

spring force.
Note: For applications that may continuously introduce occasional air bubbles

to the flow stream upstream of the device, the device may be mounted

upside down to prevent the bubbles from becoming trapped in the differential

pressure sensor ports.

If the device has been installed upside down, avoid using the bleed

screws as water may leak into the electronics housing causing

permanent damage that is not covered under warranty!

PLUMBING

Your controller is shipped with plastic plugs fitted in the port openings.

To lessen the chance of contaminating the flow stream do not remove

these plugs until you are ready to install the device.

Make sure that the water will flow in the direction indicated by the flow arrow.
Standard L-Series Meters and LC-Series Controllers have female inlet and outlet

port connections. Welded VCR and other specialty fittings may have male ports.
The inlet and outlet port sizes (process connections) for different flow ranges

are shown on pages 51-55.
Instruments with M5 (10-32) ports have O-ring face seals and require no

sealant or tape. Do not use tape with welded or o-ring fittings.
For non M5 (10-32) ports use thread sealing Teflon® tape to prevent leakage

around the port threads.

Do not wrap the first two threads. This will minimize the possibility of getting

tape into the flow stream and flow body.

Do not use pipe dopes or sealants on the process connections as these

compounds can cause permanent damage to the controller should they get

into the flow stream.

When changing fittings, carefully clean any tape or debris from the port threads.
We recommend that a 20 micron filter be installed upstream of meter/

controllers with full scale ranges of 100 ccm or less and a 40 micron filter be

installed upstream of meter/controllers with full scale ranges above 100 ccm.

Avoiding long runs of small diameter tubing upstream or downstream

of the device will reduce water hammer.