beautypg.com

Alicat LC Series Liquid Flow Controller User Manual

Page 4

background image

4

5. Now you can safely tare (zero) a meter (page 19) using the

tare button on the device, or in the case of a water flow

controller, command the zero set-point (page 28) which will

auto-tare the unit.

► Smooth and consistent source pressure is highly preferable, as any

pulsing or fluctuation in water pressure will hydraulically couple to

the differential pressure sensor, and the resultant flow reading will

jump analogously.

► Please note that, depending on model and flow range, water flow

meters have an inherent internal pressure drop higher than that

of standard gas flow meters of similar flow range. The minimum

water device pressure drop is 2 psig, going all the way to 20 psig

drop in certain models. Your source input pressure must meet or

exceed this specification, in addition to any back pressure you

may have on the downstream end of your process. Please see

pages 51 - 54 of the manual or check our online specifications

chart for model specific details.

Please contact Alicat if you have any questions. We will be happy to help.

E-mail: [email protected]

Tel: 520-290-6060 or 888-290-6060

Notes for Using Alicat Water Flow Devices continued

This manual is related to the following products: