Alicat LC Series Liquid Flow Controller User Manual
Page 26
26
MAIN
This mode defaults on power up, with fl ow
as the primary displayed parameter.
The following parameters are displayed in
the Main mode.
Line Pressure: Only on units equipped with
the opti onal pressure sensor, pressing
the PSIG butt on will move the line gauge
pressure to the primary display.
Line Temperature: Pressing the °C butt on
will move the process temperature to the
primary display.
Set Point: The set-point (SETPT)is shown
in the upper right of the display.
For informati on on changing the set-point see SETPT SOURCE, page 28.
Volume: The volume fl ow rate is defaulted on the primary display. Press the
unit of measure butt on (ccm in the example above) to return the volume fl ow
rate to the primary display.
MENU: Pressing MENU switches the screen to the Select Menu display.
#C
+21.50
+0.00
CCM
MENU
SETPT
0.00
PSIG
+13.60
Flashing Error Message: An error message ( VOV = volumetric
overrange, POV = pressure overrange, TOV = temperature overrange)
fl ashes when a measured parameter exceeds the range of the sensor.
When any item fl ashes, neither the fl ashing parameter nor the fl ow
measurement is accurate. Reducing the value of the fl ashing parameter to within
specifi ed limits will return the unit to normal operati on and accuracy.
If the unit does not return to normal operati on contact Alicat.
CCM
+0.00