Select menu – Alicat LC Series Liquid Flow Controller User Manual
Page 20
20
SELECT MENU
From Select Menu you can interact with your RS-232 / RS-485 setti ngs or read
manufacturer’s data.
Press the butt on next to the desired operati on to bring that functi on to the screen.
An explanati on for each screen can be found on the following pages.
RS232
COMM
MFG
DATA
MAIN
MISC
Miscellaneous
Communicati on Select
Main
Manufacturer Data
#C
+21.50
+0.000
CCM
MAIN
TARE V
CCM
PSIA
+13.60
+
.
MODEL
INFO
BACK
MAIN
A L I C A T S C I E N T I F I C
Ph 520-290-6060
Fax 520-290-0109
PRESS
AVG
BACK
LCD
CONTRAST
MAIN
ZERO
BAND
FLOW
AVG
BACK
MAIN
UNIT ID
A
BAUD
19200
Select Menu
Volume: The volume fl ow rate is defaulted on the primary display. Press the
unit of measure butt on (ccm in the example above) to return the volume fl ow
rate to the primary display.
MENU: Pressing MENU switches the screen to the Select Menu display.
Flashing Error Message: An error message ( VOV = volumetric
overrange, POV = pressure overrange, TOV = temperature overrange)
fl ashes when a measured parameter exceeds the range of the sensor.
When any item fl ashes, neither the fl ashing parameter nor the fl ow
measurement is accurate. Reducing the value of the fl ashing parameter to within
specifi ed limits will return the unit to normal operati on and accuracy.
If the unit does not return to normal operati on contact Alicat.