beautypg.com

Alicat LC Series Liquid Flow Controller User Manual

Page 54

background image

54

Technical Data for Alicat LC & LCR Water Flow Controllers

0 – 5 ccm Full Scale through 0 – 5 lpm Full Scale

Alicat LC and LCR-Series water flow controllers display both water flow and water

temperature. An optional pressure sensor is also available.
Standard Specifications (Contact Alicat for available options.)

Range Specific Specifications

Performance

LC Controllers

LCR Controllers

Available Ranges

5 ccm to 500 ccm

1 lpm to 5 lpm

Accuracy at calibration conditions

after tare

± 2% Full Scale

Repeatability

± 2% Full Scale

Zero Shift and Span Shift

0.02% Full Scale / ºCelsius / Atm

Operating Range / Turndown Ratio

2% to 100% Full Scale / 50:1 Turndown

Maximum Flow Rate

102.4% FS Controllable

Typical Response Time

100 ms (Adjustable)

Warm-up Time

< 1 Second

Operating Conditions

LC Controllers

LCR Controllers

Operating Temperature

+10 to +50 ºCelsius

Maximum Pressure

150 psig

Mounting Attitude Sensitivity

None

Mount with valve cylinder vertical & upright

Valve Type

Normally Closed

Ingress Protection

IP40

Wetted Materials

LC, LCR: 430FR Stainless Steel, 302 & 303 Stainless Steel, 316L

Stainless Steel, Viton®, FFKM.

LCR Controllers Add 416 Stainless Steel, Delrin®.

If your application demands a different material, please contact Alicat.

Communication / Power

LC Controllers

LCR Controllers

Monochrome LCD or Color TFT

Display with integrated touchpad

Simultaneously displays Flow and Temperature

(optional Pressure sensor available)

Digital Output Signal

1

Options

RS-232 Serial / RS-485 Serial / PROFIBUS

3

Analog Output Signal

2

Options

0-5 Vdc / 1-5 Vdc / 0-10 Vdc / 4-20 mA

Optional Secondary Analog Output

Signal

2

0-5 Vdc / 1-5 Vdc / 0-10 Vdc / 4-20 mA

Electrical Connection Options

8 Pin Mini-DIN / 9-pin D-sub (DB9) / 15-pin D-sub (DB15) / 6 pin locking

Supply Voltage

12-30 Vdc (15-30 Vdc for 4-20 mA outputs)

24-30 Vdc

Supply Current

0.250 Amp

0.750 Amp

1. The

Digital Output Signal communicates Flow and Temperature (optional Pressure sensor available)

2. The

Analog Output Signal and Optional Secondary Analog Output Signal communicate your choice of

Flow or Temperature

3. If selecting

PROFIBUS, no analog signal is available. PROFIBUS units do not have the display. See

PROFIBUS specifications for PROFIBUS supply voltages and currents. (

www.alicat.com/profibus)

Full Scale Flow

Pressure Drop

1

at

FS Flow (PSID)

Mechanical

Dimensions

Process Connections

2

LC Controllers

5 ccm to 50 ccm

5.0

4.6”H x 3.6”W x 1.1”D

1/8” NPT Female

100 ccm

5.0

200 ccm to 500

ccm

7.5

LCR Controllers

1 lpm

4.0

5.7”H x 7.7”W x 2.3”D

1/4”NPT Female

2 lpm

4.0

5 lpm

10.0

1. Lower Pressure Drops Available, please contact Alicat.

2. Compatible with Beswick®, Swagelok® tube, Parker®, face seal, push connect and compression adapter fittings. VCR and

SAE connections upon request.