OUTPUT SIGNALS
RS-232 / RS-485 Digital Output Signal
To use the RS-232 or RS-485 output signal, it is necessary to connect the RS-232

/ RS-485 Output Signal (Pin 5), the RS-232 / RS-485 Input Signal (Pin 3), and

Ground (Pin 8) to your computer serial port as shown on page 8. (See page 32

for details on accessing RS-232 / RS-485 output.)
Standard Voltage (0-5 Vdc) Output Signal
Meters/controllers equipped with a 0-5 Vdc (opti onal 0-10 Vdc) will have this

output signal available on Pin 6. This output is generally available in additi on to

other opti onally ordered outputs. This voltage is usually in the range of 0.010

Vdc for zero fl ow and 5.0 Vdc for full-scale fl ow. The output voltage is linear

over the enti re range. Ground for this signal is common on Pin 8.
Opti onal 0-10 Vdc Output Signal
If your meter/controller was ordered with a 0-10 Vdc output signal, it will be

available on Pin 6. (See the Calibrati on Data Sheet that shipped with your meter

to determine which output signals were ordered.) This voltage is usually in the

range of 0.010 Vdc for zero fl ow and 10.0 Vdc for full-scale fl ow. The output

voltage is linear over the enti re range. Ground for this signal is common on Pin 8.
Opti onal Current (4-20 mA) Output Signal
If your meter/controller was ordered with a 4-20 mA current output signal, it

will be available on Pin 1. (See the Calibrati on Data Sheet that shipped with

your meter to determine which output signals were ordered.) The current

signal is 4 mA at 0 fl ow and 20 mA at the meter’s full scale fl ow. The output

current is linear over the enti re range. Ground for this signal is common on Pin

8. (Current output units require 15-30Vdc power.)
Opti onal 2nd Analog Output Signal
You may specify an opti onal 2nd analog output on Pin 2 at ti me of order. (See

the Calibrati on Data Sheet that shipped with your meter to determine which

output signals were ordered.) This output may be a 0-5 Vdc, 0-10 Vdc, or

4-20 mA analog signal that can represent any measured parameter. With this

opti onal output, a meter could output the volume fl ow rate (0-5 Vdc on pin 6)

and the line temperature (0-5 Vdc on pin 2).

If your device is equipped with the opti onal six pin industrial connector,

please contact Alicat.