beautypg.com

Calibrator interface and operation, 000ma – Martel Electronics LC-110/H User Manual

Page 7

background image

5

2. Calibrator Interface and Operation

MARTEL ELECTRONICS

LC-110H

Precision Loop

Calibrator w/HART™

100%

25%

0%

MENU

EXIT

Source

0.00%

4.000mA

Simultaneous % Display
4mA = 0.00%
20mA = 100.00%

Cursor

100% Key: Sets Output to 20mA

25% Key: Steps 4, 8, 12, 16, 20mA

0% Key: Sets Output to 4mA

Digital Knob Sets Output Value
Press to Move Cursor

Menu
Press to Enter Setup Menu
Press Again to Exit

Step / Ramp Enable

Backlight

On / Off

Mode

Figure 1

Figure 1 is typical of the main display for all operating modes. It
shows the present measurement or output value. The individual
operating modes are described below.
Use the MENU/EXIT key to display the main menu described in
section 3 to select the primary operating mode of the calibrator, to
access the calibrator setup functions, or to enter HART mode.
In output mode:

• Press the knob to move the decade cursor.

• Rotate the knob to increment or decrement the output in steps

indicated by the selected decade.

• Use the 3 percent keys to directly set the output to preset

values.

• Use the step/ramp key to select and stop these advanced

modes which are described in section 5.1.