4 milliamp measure with 24v power, 5 volts measure, 000ma – Martel Electronics LC-110/H User Manual
Page 11
9
2.4 Milliamp Measure with 24V power
In the mA Measure with 24V mode, the calibrator outputs 24VDC
while displaying the loop current. The mode is useful for powering a
transmitter without the need for a separate power supply.
Figure 5 shows the main display and typical connections for this
mode.
MARTEL ELECTRONICS
LC-110H
Precision Loop
Calibrator w/HART™
100%
25%
0%
MENU
EXIT
Measure 24V
0.00%
4.000mA
Figure 5
• Press MENU/EXIT to enter the main menu described in section 3.
• The text ‘250
Ω’ text is shown in the lower right corner when the
HART resistor has been enabled on the Setup Menu, section 4.4.
2.5 Volts Measure
In the Volts Measure mode, the calibrator displays the loop voltage.
Figure 6 shows the main display and typical connections for this
mode.
• Press MENU/EXIT to enter the main menu described in section 3.
- 3001 (94 pages)
- BetaGauge 301 (40 pages)
- BetaGauge 311A-Ex (32 pages)
- BetaGauge 311A (48 pages)
- BetaGauge 330 (48 pages)
- BetaGauge 3II (197 pages)
- BetaGauge PI (20 pages)
- BetaGauge PIR (24 pages)
- BetaLOG-PI (28 pages)
- BetaLOG-TI (16 pages)
- BetaProbe TI/TI+ (9 pages)
- DPC-30 (12 pages)
- DPC300A (48 pages)
- DMC1410 (108 pages)
- IVC-222HPII (12 pages)
- BETA LDT-500 (2 pages)
- M2000 (28 pages)
- MC1010 (46 pages)
- MC1210 (56 pages)
- TC100 (16 pages)
- T150 (8 pages)
- T140 Series (12 pages)
- PTC8010 (36 pages)
- PSC4010 (36 pages)
- MS-420 (1 page)
- MECP500 (2 pages)
- MECP2000 (2 pages)
- MECP10K (4 pages)
- MECP100 (2 pages)
- T130 (2 pages)
- T120 (2 pages)
- MC1200 (50 pages)
- MC1000 (28 pages)
- MasterCAL (140 pages)
- LC100 (20 pages)
- DMC1400 (106 pages)
- 320 (60 pages)
- 230 (106 pages)
- 130 (85 pages)
- QM100V (1 page)
- QM100V-40 (1 page)
- QM110T (2 pages)
- QM120ET (2 pages)
- QM130M (2 pages)