Inrush current measurement, Steps for measurement, Timing of data recording – KYORITSU 6310 Quick Manual User Manual
Page 33

−
31 −
Inrush Current Measurement
KEW6310
Inrush Current Measurement
Steps for measurement
Ensuring your safety
↓
Preparation for measurement
↓
Setting
↓
Wiring
↓
Inrush Current Measurement
Range
* Readings are displayed right after the inrush current measurement starts.
Timing of data recording
Measured data will be saved when an event occur or at the preset interval during measurement.
Recording at event occurrence
Recording at every interval
File ID:
6310-09
File ID:
6310-15
Inst value : Avg of 100 data (@50Hz) obtained 1 sec before the preset interval comes (rms)
Avg value : Avg of rms values obtained in the preset inst interval
Max value : Max rms values obtained in the preset inst interval
Min value : Min rms values obtained in the preset inst interval
KEW6310
Range
Measurement setting
Save Setting
Inrush Current Measurement Recording method
Interval
Recording start
Clamp sensor
Recording termination
A Range
Destination to save data
Reference current
Destination to save screen shot
Filter
Threshold value
Hysteresis
Trigger point
Setting
Example
Reference current
100A
Threshold value
110%
Hysteresis
1%
Trigger point
Before: 100
After: 100
Setting
Example Interval
30min