29 −
Transient measurement
KEW6310
Transient measurement
Steps for measurement
Ensuring your safety
↓
Preparation for measurement
↓
Setting
↓
Wiring
↓
Transient Measurement
Range
* Readings are displayed right after the recording of Transient measurement starts.
Timing of data recording
Measured data will be saved when an event occur or at the preset interval during measurement.
Recording at event occurrence
Recording at every interval
File ID:
6310-08
File ID:
6310-14
Inst value : max value of 10,000 data obtained at 100us 1sec before the preset interval comes
Avg value : Avg of inst values obtained in the preset inst interval
Max value : Max inst values obtained in the preset inst interval
Min value : Min inst values obtained in the preset inst interval
KEW6310
Range
Measurement setting
Save Setting
Transient measurement
Recording method
Interval
Recording start
V Range
Recording termination
Threshold value
Destination to save data
Hysteresis
Destination to save screen shot
Trigger point
Setting
Example
Threshold Value
110%
Hysteresis
1%
Trigger point
Before: 100
After: 100
Setting
Example Interval
30min
Interval(30min)
Interval
(reference
Voltage)
Past100
date pts
201date pts
recorded
Trigger
Start
Peak(max)
Ave, max and min* values within
each interval are recorded.
Past100
date pts
Threshold
110%
100%