Charging the rechargeable Ni-MH batteries
Following message to prompt battery charge appears on the LCD automatically when battery level is
40% or less at starting the instrument. Press the
Cursor
Keys and
ENTER
Key according to
the instructions displayed on the LCD.
・Install rechargeable batteries (Ni-MH)
・Slide the Selector switch to the right (set to “RE-CHARGEABLE” position)
・Connect the AC Power cord and power on the instrument.
* Refer to “(4.2.4.) Other Setting” in the full instruction manual to initiate a battery charge anytime it is necessary.
Battery charge doesn't initiate only by installing rechargeable batteries and connecting an AC power cord.
Above operation is required to start a battery charge.
How to install batteries:
Install batteries in correct
polarity as marked inside.
Battery power is consumed even if the instrument is being off. Remove all the batteries if the
instrument is to be stored and will not be in use for a long period.
KEW6310
KEW6310
Charging the rechargeable Ni-MH batteries
−
8 −
“
Charge batteries?”
No
Return to normal screen.
(Batteries aren't charged.)
Yes
Proceed to next screen
“
Rechargeable batteries are installed?”
No
Return to normal screen.
(Batteries aren't charged.)
Yes
Battery charge starts; return to