beautypg.com

Basic setting, Setting item details of setting, Wiring configuration – KYORITSU 6310 Quick Manual User Manual

Page 13

background image

KEW6310

Basic Setting 

KEW6310

11 −

Basic Setting

Setting item

Details of Setting

Wiring

1P2W×1

1P2W×2

1P2W×3

1P2W×4

1P3W×1

1P3W×2

1P3W×1+2A

3P3W×1

3P3W×2

3P3W×1+2A

3P3W3A

3P4W×1

3P4W×1+1A

4A

Voltage Range

150V╱

300V╱600V╱1000V

VT Ratio

0.01~9999.99(1.00

Clamp / Current Range

8128:1╱5╱10╱20╱50A╱AUTO

8127:10╱20╱50╱100A╱AUTO

8126:20╱50╱100╱200A╱AUTO
8125:50╱100╱200╱500A╱AUTO

8124:100╱200╱500╱1000A╱AUTO

8129:300╱1000╱3000A

8141:

8142:

100╱500mA╱1A╱AUTO

8143:

8146:

8147:

500mA╱1╱5╱10A╱AUTO

8148:

CT Ratio

0.01~9999.99( 1.00

Filter

8141╱42╱43╱46╱47╱48:ON╱OFF
8128╱27╱26╱25╱24╱29: --------

DC V

50mV╱500mV╱ 5V

Frequency

50Hz╱60Hz

            

* Default values are highlighted in gray

            

* Leakage Clamp sensors cannot be used for power measurements but can be used on

wiring configurations: ⑦, ⑩, ⑬ and ⓪.

Wiring Configuration





Power Clamp sensor

Leakage Clamp sensor