KROHNE BM 90 EN User Manual

Page 39

background image

39

Points

’A’

’b’

Number

% Head or Level

% of Flow of volume

1

0

0.0

2

10

0.0

3

20

7.2

4

30

16.1

5

40

27.3

6

50

37.5

7

60

48.5

8

70

59.5

9

80

70.5

10

90

80.0

11

100

89.5

12

110

100.0

13

255

-

14

255

-

15

255

-

16

255

-

110

100

90

80

70

60

50

40

30

20

10

0

Vessel

(or

channel)

symmetrical

about

c/1

Transducer

%

of

level

Point

Head (cm)

% Head

’A’ Value

Flow litres/sec

% Flow

’b’ Value

1.00

0.0 0.0

0.00

0.0 0.0 0

2.00

2.5 8.3

8.00

0.4

1.1

11

3.00

5.0

16.70

17.00

1.7

4.4

44

4.00

7.5

25.0

25.00

3.7

9.5

95

5.00

10.0

33.30

33.00

6.3

16.2

162

6.00

15.0 50.0

50.00

12.70

32.5

325

7.00

20 0

66.60

67.00

20.30

52.0 520

8.00

25.0 83.20

83.00

29.30

75.0 750

9.00

30.0 100.0

100.00

39.0 100.0

1000

10.00

Not used

Not used

255.00

Not used

Not used

.0

11.00

Not used

Not used

255.00

Not used

Not used

.0

12.00

Not used

Not used

255.00

Not used

Not used

.0

13.00

Not used

Not used

255.00

Not used

Not used

.0

14.00

Not used

Not used

255.00

Not used

Not used

.0

15.00

Not used

Not used

255.00

Not used

Not used

.0

16.00

Not used

Not used

255.00

Not used

Not used

.0

NOTE:

1. Points 10 to 16 not used - leave at factory

default value.

2. ’A’ values must be whole numbers, no deci

mals allowed.

3. ’b’ values must be entered as the tabulated

value, the decimal will be automatically allocated.

Now continue programming the instrument as follows:

Programme the instrument exactly as section 4.5 on Page 31,

except:

Change

Pr.45 from 2 to 4, which denotes

"Special Flow Device".

Go to

Pr.44 and proceed as follows:

Press

Pr.44 to display Pr.44:

Press

’DSP’

to show

Press

’ENT’

to show ’A1’

Press

’DSP’

to show value of ’A1’ (default = 255)

Key in the value ’O’ from Table 1 and press ’ENT’

Press

’DSP’

to show ’A1’ again

Press ’

s’

to show ’A2’

Press

’DSP’

to show value of ’A2’ (default = 255)

Key in the value ’8’ from Table 1 and press ’ENT’

Continue for all points which you need to use

(up to ’A16)’

Any points not used must be left at the default

of 255

Press

’DSP’

to display the last ’A’ number used, then

Press

t’

several times to return to ’A1’

Press

#

to show ’b1’

Press

’DSP’ to show value of ’b1’ (default = .0)

Key in the value ’0’ from Table 1 and press ’ENT’

Press

’DSP’ to show ’b1’ again

Press

s’

to show ’b2’

Press

’DSP’ to show value of ’b2’ (default = .0)

Key in the value ’11’ from Table 1 (accepted as 1.1)

Continue for all points which you need to use (up to 16)

Any points not used must be left at the default of .0

Press

’TEST’ then ’DSP’ to show Pr.44

Leave the linearisation part of the programme by dis

playing any other parameter, or go into ’RUN’ mode.

Example: Flow - Special Flume Mapping

Use Example 4.5 on Page 31, but substitute a special ’U’ throat flume with maximum flow

39 litres/second at 30 cm/hd.

First, create the following table from the relationship of head and flow which must be given forthe special flume.