Transducer mounting – KROHNE BM 90 EN User Manual

Page 11

background image

11

2.2. Transducer Mounting

The transducer can be supplied as ’standard’ or mounted in a Teflon faced flange for applications requiring chemical

compatibility. The figure below shows the dimensions:

20

170

Isolation Kit

Fit as shown

20

21

72

100

10m cable
6.9 dia

50

M20 x 1.5

95

Dia to suit flange selected

No. & size of bolts
to suit flange selected

Bolt hole PCD to suit

flange selected

DIA

DIA

DIA

Standard

Flanged

An isolation kit is provided with each transducer to min-

imise any ringing transmitted through the mounting struc

ture.

The transducer must be mounted perpendicular to the

monitored surface and, ideally, at least 0.5 metres above

i

t

.

The transducer has a 10 inclusive conical beam angle

at 3dB and must be mounted with a clear unobstructed

sight of the liquid to be measured over the complete

measurement range.

The transducer is provided with integral cable which can

be extended up to 300 metres using a suitable junction

box and RG62AU cable. The temperature compensated

transducer requires an additional single core screen

extension.

The extended cable should then be terminated directly

into the instrument.

Transducer cables and temperature compensation

cables can be run together but should be separated from

power cables by at least 150mm and preferably installed

in their own earthed steel conduit.