KEEP TRANSDUCER

PERPENDICULAR

TO LIQUID.

KEEP TRANSDUCERS AND

TEMPERATURE COMPENSATION

PROBES OUT OF DIRECT

SUNLIGHT

DO NOT AIM THROUGH

HOLES IN THE TANK

.

AVOID ROUGH EDGES

IN STANDPIPES

.

AVOID INFLOWS OR

OTHER OBSTRUCTIONS

.

CAUTION: AVOID THE FOLLOWING TRANSDUCER INSTALLATION FAULTS

Standpipe Installations

In many applications access to a vessel must be made

via a standpipe. However, it is necessary to observe

some basic rules when fitting transducers into stand-

pipes.

BLANKING:

Parameter 5 should always be set at

least 150 mm longer than the length of

the standpipe.

STANDPIPE

should be in accordance with the follo-

wing table

DIMENSIONS:

Flange size and minimum

Maximum length

bore of Standpipe

of Standpipe

3"

( 75mm)

3 0 0 m m

4"

(100mm)

3 0 0 m m

6"

(150mm)

4 0 0 m m

8"

(200mm)

6 0 0 m m

12"

(300mm)

6 0 0 m m

e.g. : Using a 4" flanged transducer would require the

standpipe length to be no more than 300mm and Pr.5 set

at 450 mm minimum.

The inside of the pipe and joint with vessel top must

be clean and free of any obstructions, seams or

welds.

Isolation kit

dia 95 (3.74")

M20 X 1.5

22 (0.86")

155 (6.10")

P V C
flange