B.67 [source:]voltage:protection:tripped? query, B.68 [source:]function:mode? query, B.69 status:operation:condition query – KEPCO ABC-DM SERIES User Manual
Page 86: B.70 status:opereration:enable command, B.71 status:opereration:enable? query, B.67, B.68, B.69, B.70, B.71
B-16
ABC 040104
B.67
[SOURce:]VOLTage:PROTection:TRIPped?
QUERY
VOLT:PROT:TRIP?
Syntax:
Short Form: [SOUR:]VOLT:PROT:TRIP? Long Form: [SOURce:]VOLTage:PROTection:TRIPped?
Return Value: 0 or 1
Description: Identifies whether overvoltage condition was detected. Returns 0 if overvoltage protection
tripped; 1 if overvoltage protection not tripped. Related Commands: VOLT:PROT, VOLT:PROT:CLE.
(See example, Figure B-6.)
B.68
[SOURce:]FUNCtion:MODE?
QUERY
FUNC:MODE?
Syntax:
Short Form: FUNC:MODE?
Long Form: [SOURce:]FUNCtion:MODE?
Return Value: VOLT or CURR
Description: Identifies the operating mode of the power supply. VOLT = Constant Voltage mode (CV).
CURR = Constant Current mode (CC). (See example, Figure B-1.)
B.69
STATus:OPERation:CONDition
QUERY
STAT:OPER:COND?
Syntax:
Short Form: STAT:OPER:COND?
Long Form: STATus:OPERation:CONDition?
Return Value:
Description: Returns the value of the Operation Condition Register (see Table B-2). The Operation Condition
Register contains unlatched real-time information about the operating conditions of the power supply.
Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inactive, false). (See
example, Figure B-7.)
B.70
STATus:OPEReration:ENABle
COMMAND
STAT:OPER:ENAB
Syntax:
Short Form:
STAT:OPER:ENAB
Long Form:
STATus:OPERation:ENABle
Description: Sets Operation Enable Register. The Operation Enable Register is a mask for enabling specific bits
in the Operation Event Register which will cause the operation summary bit (bit 7) of the Status Byte
register to be set Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inac-
tive, false). The operation summary bit is the logical OR of all the enabled bits in the Operation Event
register. (See example, Figure B-7.)
B.71
STATus:OPEReration:ENABle?
QUERY
STAT:OPER:ENAB?
Syntax:
Short Form: STAT:OPER:ENAB?
Long Form: STATus:OPERation:ENABle?
Return Value:
Description: Reads Operation Enable Register (see Table B-2). Returns value of Operation Enable Register
bits. Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inactive, false).
(See example, Figure B-7.)
TABLE B-2. OPERATION CONDITION REGISTER, OPERATION ENABLE REGISTER,
AND OPERATION EVENT REGISTER BITS
CONDITION
NU
CC
NU
CV
NU
WTG
NU
CAL
BIT
15-11
10
9
8
7 - 6
5
4 - 1
0
VALUE
32,768 -
2048
1024
512
256
128 - 64
32
16 -2
1
CAL - INTERFACE COMPUTING NEW CALIBRATION
CONSTANTS
CC -
POWER SUPPLY IN CONSTANT CURRENT MODE
CV -
POWER SUPPLY IN CONSTANT VOLTAGE MODE
NU -
NOT USED
WTG WAIT FOR TRIGGER (SET BY INIT SUBSYSTEM)