beautypg.com

B.65 [source:]voltage:protection[:level]? query, Figure b-6. programming voltage, B.66 [source:]voltage:protection:clear command – KEPCO ABC-DM SERIES User Manual

Page 85: B.65, B.66, Volt:prot, Volt:prot:cle

background image

ABC 040104

B-15

protection level can be programmed independently of the output voltage level. Setting voltage protec-
tion to a value lower than the output voltage causes the voltage protection mechanism to trigger
immediately. (See example, Figure B-6.)

B.65

[SOURce:]VOLTage:PROTection[:LEVel]?

QUERY

VOLT:PROT?

Syntax:

Short Form: [SOUR:]VOLT:PROT[:LEV] MIN, MAX
Long Form: [SOURce:]VOLTage:PROTection[:LEVel] MIN, MAX
Return Value: = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1

Description: Identifies overvoltage protection setting, maximum allowable overvoltage protection, or mini-

mum overvoltage protection. VOLT:PROT? returns value set by VOLT:PROT. VOLT:PROT? MAX
returns maximum voltage protection value (see Table 1-2 ); this value is determined at the factory and
cannot be changed by the user. VOLT:PROT? MIN returns the minimum voltage protection value
(always 0). (See example, Figure B-6.)

FIGURE B-6. PROGRAMMING VOLTAGE

B.66

[SOURce:]VOLTage:PROTection:CLEar

COMMAND

VOLT:PROT:CLE

Syntax:

Short Form: [SOUR:]VOLT:PROT:CLE Long Form: [SOURce:]VOLTage:PROTection:CLEar

Description: Clears overvoltage condition. Output programmed to 0V; current limit programmed to small value

(approximately 1% of full scale current rating). Overvoltage Protection Tripped is reset so that
VOLT:PROT:TRIP? returns 0. Related Commands: VOLT:PROT, VOLT:PROT:TRIP?. (See example,
Figure B-6.)

NOTES: 1. The power supply is assumed to be operating in constant voltage (CV) mode.

2 Examples below are intended only to illustrate command functions. Refer to PAR. 3.3.8.1 for pro-

gramming techniques to optimize performance

VOLT 21; CURR 1.1

Power supply programmed to 21V, 1.1A.

VOLT 2.157E1

Power supply programmed to go to 21.57V.

VOLT?

Returns 2.157E1, indicating programmed voltage = 2.157V.

VOLT? MAX

For ABC 135-1DM, returns 135 (135V).

VOLT? MIN

Returns 0 (minimum allowable voltage).

VOLT:PROT 2.365E+1

Voltage protection set to 23.65V.

VOLT?

Returns 21.57, indicating programmed voltage value = 21.57V.

VOLT:PROT?

Returns 2.365+E1, (programmed overvoltage protection = 23.65V

VOLT:PROT?MAX

For ABC 10-10DM, returns 11.

VOLT:PROT:TRIP?

0 indicating overvoltage protection not tripped.

--- OVERVOLTAGE CONDITION OCCURS.

VOLT:PROT:TRIP?

1 indicating overvoltage protection tripped.

VOLT:PROT:CLE

Overvoltage condition cleared.

VOLT:PROT:TRIP?

0 indicating overvoltage condition not tripped.

VOLT?

Returns 0, programmed voltage value = 0V.

VOLT 21; CURR 1.1

Power supply programmed to go to 21V, 1.1A.

VOLT?

Returns 21.

VOLT:LIM:HIGH 5

Voltage limit set to 5V.

VOLT:LIM:HIGH?

Returns 5.

VOLT 33

Power supply programmed to 5V.

SYST:ERR?

Error message -301 “Value bigger than limit.”

VOLT?

Returns 5.