4 changing the calibration password, Table 4-1. factory default calibration passwords, Changing the calibration password -3 – KEPCO ABC-DM SERIES User Manual
Page 61: Factory default calibration passwords -3, 4 to c
ABC 040104
4-3
4. LCD will flash
Iout F.S.adj. **WAIT**
then
Iout F.S. adj < >
. Monitor DVM and
use
or
keys to increase or decrease output voltage as necessary to get DVM reading
as close as possible to value listed below (without exceeding value listed). If the
"
and
#
keys do not provide the required adjustment range, press the 1 (increase) or 3 (decrease)
key for additional range.
–
ABC 10-10DM:
1.000V
–
ABC 15-7DM:
0.700V
–
ABC 25-4DM:
1.000V
–
ABC 36-3DM:
0.750V
–
ABC 60-2DM:
2.000V
–
ABC 125-1DM:
1.000V
5. Press ENTER when complete.
4.4
CHANGING THE CALIBRATION PASSWORD
The 4-digit password is required for to enter calibration mode. The factory default passwords
are listed in Table 4-1.
1. If the unit is not in command entry status , press CLEAR or RESET to put the unit in com-
mand entry status.
2. Press MENU key until LCD reads
Change Passw. OldPass
.
3. Enter the old password and press ENTER. Then enter the new 4-digit Password and press
ENTER.
4. The new password is accepted and the LCD shows the next Menu screen
GoTo Previous
CALIB? 1 Yes
.
5. Press CLEAR to exit the Menu, press 1 to restore the previous calibration values (see PAR.
4.5), or MENU for additional menu screens.
TABLE 4-1. FACTORY DEFAULT CALIBRATION PASSWORDS
MODEL
PASSWORD
ABC 10-10DM
1010
ABC 15-7DM
1507
ABC 25-4DM
2504
ABC 36-3DM
3603
ABC 60-2DM
6002
ABC 125-1DM
1251