KANOMAX 3800 Handheld Condensation Particle Counter User Manual

Page 40

background image

33

Symptom Possible

Cause

Maintenance/Solution

LASER ERR】

Message

Possible laser damage.

Send the instrument to our factory for
repair.

PUMP ERR】

Message

Possible pump damage.

Send the instrument to our factory for
repair.

TEMP ERR】

Message

Possible Peltier damage.

Send the instrument to our factory for
repair.

ALCOHOL ERR】

Message

Alcohol shortage.

Charge the alcohol cartridge with alcohol.

Load is applied to the inlet.

Confirm that there is no load applied to the
inlet.

FLOW ERR】

Message

Alcohol cartridge or storage cap is
not properly inserted.

Properly insert and turn the cartridge until
it cannot be turned any further.

COUNT OVER】

Message

Measuring particle concentration
is too high.

Measurement is possible, however, the
optical system may become contaminated.
Make sure to attach the filter to the inlet
for cleaning after use.

COM ERR】

Message

Communication cable to the PC is
disconnected.

Confirm the connection with the PC.

Communication cable to the
printer is disconnected.

Confirm the connection with the printer.

PRINT ERR】

Message

Communication setting is
incorrect.

Communication setting of the printer must
be consistent with the CPC.
See 3.3.4.2 Printer Setup (page 25).

PUMP STOP】

Message

Instrument is tilted during
operation.

Tilt of the instrument causes the pump to
stop. Put the instrument in a horizontal
position. The pump will start operation in 5
seconds after the instrument is returned to
a horizontal position.

Power does not turn
on.

Low battery.

AC adapter is not functioning or
not connected.

Replace the batteries.

Connect the AC adapter.