2 measurement modes – KANOMAX 3800 Handheld Condensation Particle Counter User Manual
Page 20

13
3.2 Measurement Modes
There are four (4) measurement modes as listed below.
1.
After warm-up, the instrument will automatically transfer to this mode.
Particle number concentration (particles/cc) and absolute pressure (hPa) can be measured.
Measured data cannot be stored.
This mode also allows setting of the Menu and printing an instantaneous value.
2.
The measurement is performed in accordance with the measurement condition set by the CPC.
Particle number concentration (particles/cc) and absolute pressure (hPa) can be measured.
Measured data can be stored.
3.
The measurement is performed in accordance with the measurement program .set by the PC,
and written in the CPC.
Particle number concentration (particles/cc) and absolute pressure (hPa) can be measured
automatically.
Measured data can be stored.
4.
Counts up the number of particles measured during a random time interval.
The number of particles and absolute pressure (hPa) can be measured.
Measurement Start/Stop is manually operated.
Measurement data cannot be stored.
This mode is suitable for performing a measurement in an environment where the particle
number concentration is low.