beautypg.com

KANOMAX 3800 Handheld Condensation Particle Counter User Manual

Page 23

background image

16

3.2.3 Programmed Mode

This is a mode for performing automatic measurement
in accordance with the measurement program(s)
written in from the PC.
Maximum 5 programs can be set.
Select . Then, in order to activate the
programmed measurement, press and hold the
[ENTER] key for a few seconds, and 【START】 will be
indicated on the screen

(Screen-6)

.

When 【START】 is indicated, the CPC is ready for
programmed measurement.
Make sure that 【START】 is indicated. If not indicated, the programmed measurement will NOT
be performed.
If the time interval before the programmed start time is 15 minutes or more, the pump and Peltier
operation will stop, and the message 【WAITING TIME】 will be indicated on the screen.

In this mode, the pump and Peltier will start operation 10 minutes before the measurement start
time of the program(s) 1-5, and the message 【WARMING UP…】 will be displayed on the screen.
The measurement will start automatically at the exact start time.

If there is any overlap between the measurement times set by the program(s) 1-5, the program
with an earliest start will have a priority over the other overlapping program(s), and the other
overlapping program(s) with the later start time will become invalid.
For canceling (deactivating) this mode, press and hold the [ENTER] key for a few seconds, and the
automatic measurement will be cancelled. The 【START】 indication on the display will turn off.

To perform daily measurement at a programmed time, set the program date as 00/00/00.
The CPC must be powered up in order to perform the programmed measurement.

While performing a measurement, the power switch cannot be operated. To turn off the power, the
measurement must be terminated first.

For details of the items set by the programmed measurement, see 3.3.2 Viewing Program (page 20),
and Section 5 “CPC Settings” of the Software Operation Manual for how to set the programmed
measurement.

Screen-6