beautypg.com

KANOMAX 3800 Handheld Condensation Particle Counter User Manual

Page 28

background image

21

3.3.3 Data Output

To view, print and delete stored data:
On the MENU

(Screen-8)

select “4. DATA OUTPUT”, and press the [ENTER] key.

The display will switch to a screen displaying a count result of the latest stored data

(Screen-11)

.

* When there is no data, you will see a message “NOT RECORDED”, and the display will

automatically return to the MENU

(Screen-8)

.

Each item indicates:
“S-TIME” = Sampling length;,
“S-TIMES”= Number of sampling frequency
“AVG/MAX/MIN”= Average/Maximum/Minimum value of particle number concentration under
measurement.
“P AVG/P MAX/P MIN”= Average /Maximum/ Minimum value of absolute pressure under
measurement.

To view previous or next data from the displayed data, press the [UP]/[DOWN] key.
To return to the MENU

(Screen-8)

,

press the [ENTER] key

.


To view the details of the count result displayed on

Screen 11

, press the [LEFT] key and detail data

will be displayed

(Screen-12)

.

Each item indicates:
“TIME” = Sampling length
“DATA” = Particle number concentration [particles/cc]
“PRES.” = Absolute pressure at inlet [hPa]

“S”= Error status
For details of the error status, see 1.2 Operation Panel, List of Error Messages (page 3).
If the number of “S-TIMES” is 7 or more, you can move to the next page by using the [UP]/[DOWN]
key.
To return to the count result screen

(Screen-11)

, press the [LEFT] key.


To switch to the DATA UTILITIES screen

(Screen-13)

where each count result (and its detail data)

can be deleted and/or printed by using an optional printer, press the [RIGHT] key on the count
result screen

(Screen-11)

.