KANOMAX 3800 Handheld Condensation Particle Counter User Manual

Page 24

background image

17

3.2.4 Count Mode

This is a mode for counting up the number of particles.
The readings (measured value) will be updated
once each second.
Measurement will start on pressing
the [ENTER] key, and the number of particles
per second will be accumulated.

On the display, elapsed time and the
accumulated number of particles will be indicated.


By pressing the [ENTER] key, counting up of particles and elapsed time will stop, and the
measurement will be terminated.
When the [ENTER] key is pressed again, the display will return to the initial display of the
Mode.
The maximum count is 9, 999, 999. When this is exceeded, the message 【COUNT OVER】 will be
displayed on the screen.
In this mode, measured values cannot be stored.
While performing a measurement, the power switch cannot be operated. To turn off the power, the
measurement must be terminated first.

Screen-7

<COUNT> Mode

Elapsed time