List of Error Messages (In priority order)

1.【LASER ERR】---Laser failure.
2.【PUMP ERR】--- Pump failure.
3.【TEMP ERR】--- Temperature control failure
4.【ALCOHOL ERR】--- Alcohol supply failure.
5.【FLOW ERR】--- Flow control failure.
6.【COUNT OVER】--- Count exceeding 100,000

particles/cc, or in mode, when
exceeding 9,999,999.

7.【COM ERR】--- Communication error.
8.【PRINT ERR】--- Printer output failure.

List of Status Display

1.【HOLD】

Screen is in “hold” for printing.

2.【PUMP STOP】
Pump is stopped on detecting a tilt condition.
3.【ALARM】
Set alarm level is exceeded under an alarm

activated condition.

4.【MEMORY OVER】
No memory remaining for data storage.
5.【WAITING TIME】
When operating in , the instrument

is in standby during an interval.

6.【WARMING UP…】

When operating in , the instrument
is warming up to recover from standby (10
min.)


* In case of having multiple errors at the same time, one error message will be displayed on the

screen in the priority order as shown above.
On the printed data, the error status will be indicated by the error message number which
corresponds to the above. For details, see 3.3.3 Data Output (page 21).
e.g. When having the error statues 2 and 3, “S”=2 will be shown on the printed data.

If indicated as “S”=5, this means that there was a flow error.

* For measures to be taken to solve an error status, see Section 6. Troubleshooting (page 32).