- Assembling the Cartridge





Insert a clean mesh into the cartridge cylinder and confirm that it lies flat on the bottom.
Then insert the felt into the cartridge cylinder until it hits the bottom.
Reassemble the cartridge in the reverse order of disassembling.
Blow air to the entire unit when finished.

4.3 Frequency of Maintenance

The CPC requires very little maintenance. However, maintenances as suggested below should be
performed to ensure proper operation. There are some items that are recommended to be
performed each time the CPC is used. Other items should be performed according to the
frequency of use.

Item Frequency

Daily zero check

Daily (or before use if not used daily)

Charging the felt with alcohol

Daily (or before use if not used daily)

Replacing the alcohol felt

As required

(See chapter 6. “Troubleshooting”)

Returning the instrument to the

distributor for cleaning and calibration.

Once a year




Caution

Before assembling, make sure that all parts are clean.
If small debris or lint of the felt gets into the instrument, it may cause serious
problems.
Blow air onto the entire surface of the alcohol cartridge and felt to make sure
that they are free from dust and dirt.