In-Situ TROLL Link Telemetry System Operators Manual User Manual
Page 37

800-446-7488
37
www.in-situ.com
Problem
Possible Solution
Unable to connect the TROLL
Link Telemetry System to Win-
Situ 5 Plus Software in TCP/IP
mode.
• Ensure the correct IP Address is entered. See page
24.
• Ensure the correct Port Number is entered. See page
24.
• Ensure the correct Device Address is entered. See
page 24.
• Ensure the communication mode is set to ASCII. See
page 9.
• Ensure the Transmission Delay is set between 15 to
20 seconds. See page 24.
• Ensure all communication settings on the instrument
are correct. See page 8.
• Ensure the APN information is entered correctly using
the TROLL Link Setup Software. See page 24.
Unable to connect the TROLL
Link Telemetry System to Win-
Situ 5 Plus Software in dial-up
mode.
• Ensure the correct Modem number is selected. See
page 23.
• Ensure the correct phone number is entered. See
page 23.
• Ensure the correct Device Address is entered.
See page 23.
• Ensure the communication mode is set to ASCII. See
page 9.
• Ensure the Transmission Delay is set between 15 to
20 seconds. See page 23.
• Ensure all communication settings on the instrument
are correct. See page 8.
Unable to connect the TROLL
Link Telemetry System to Win-
Situ 5 Plus Software in SMS
mode.
A different mode must be selected when a direct
connection to the system is required.