In-Situ TROLL Link Telemetry System Operators Manual User Manual
Page 18
800-446-7488
18
www.in-situ.com
1. Click the Messages tab.
2. Select the appropriate destination for the SMS messages. SMS messages are only
sent to one device. Messages sent to an email address require a gateway phone
number. Contact your service provider for their gateway phone number. In-Situ
service plans provided by AT&T have a gateway phone number of 121.
• Select A phone. Enter the destination mobile phone number, and a description (a
site name or other information) if desired.
-OR-
• Select An email address. Enter a gateway phone number, and enter the
destination email address.
3. In the Retries field, enter the number of times the TROLL Link Telemetry System
resends a message (1 to 10 times) when communications fail. 3 is the
recommended number.
4. In the TimeOut (secs) field, enter the amount of time the TROLL Link Telemetry
System attempts to send a message before it is considered failed. 30 (the default
value) is the recommended value.
5. Click Add to set up individual messages.
6. Select an instrument from the Device drop-down menu.
7. Select the desired parameters for the SMS message from the Available Parameters
window. Click the right arrow button to move the parameters to the Message
Parameters window.
l
One SMS message holds up to five measured values from a single
instrument.
l
Instruments that require more than five values to be transmitted require
additional messages.