beautypg.com

Caution – HT instruments VEGA78 User Manual

Page 6

background image

VEGA78

EN - 4

1.3. DURING

USE

Please read carefully the below points:

CAUTION

Should you fail to keep to the prescribed instructions you could damage the
instrument and/or its components or endanger your safety.


 Do not touch any unused terminal when the meter is connected to the circuit on test.

 When measuring current, other currents located near the leads may affect the

measuring accuracy.

 When measuring current, always position the wire in the very middle of the jaws in

order to obtain the highest accuracy.

 A measured value remains constant if the "HOLD" function is active. Should you notice

that the measured value remains unchanged, disable the “HOLD” function.


1.4. AFTER

USE

 After use, turn off the instrument by pressing ON/OFF key for a few seconds.

 If you expect not to use the instrument for a long time please keep to the storage

instructions described at § 3.4.


1.5. MEASURING

(OVERVOLTAGE) CATEGORIES DEFINITIONS

IEC / EN61010-1: Safety requirements for electrical equipment for measurement, control
and laboratory use, Part 1: General requirements, gives a definition of measuring
category, usually called overvoltage category. § 6.7.4: Measuring circuits:

(OMISSIS)

Circuits are divided into the following measurement categories:

Measurement category IV is for measurements performed at the source of the low-

voltage installation.

Examples are electricity meters and measurements on primary overcurrent protection
devices and ripple control units.

Measurement category III is for measurements performed in the building installation.

Examples are measurements on distribution boards, circuit breakers, wiring, including
cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and
equipment for industrial use and some other equipment, for example, stationary
motors with permanent connection to fixed installation.

Measurement category II is for measurements performed on circuits directly

connected to the low voltage installation.

Examples are measurements on household appliances, portable tools and similar
equipment.

Measurement category I is for measurements performed on circuits not directly

connected to MAINS.

Examples are measurements on circuits not derived from MAINS, and specially
protected (internal) MAINS-derived circuits. In the latter case, transient stresses are
variable; for that reason, the norm requires that the transient withstand capability of the
equipment is made known to the user.