VEGA78

EN - 28

5.2.6. HARM analysis screens
Starting from any page of numerical values it’s possible to enter harmonics numerical
values and histogram graphics screens section for voltage and current input signals by
pressing F2 key (or touch HARM at display). By pressing cyclically F1 key the below
screens are shown by meter:
 Harmonics values of V1, V2, V3 voltages and neutral voltage Vn (for three phase 4-

wire system), currents I1, I2, I3 and neutral current In (for three phase 4-wire
system)with THD% values both with histogram graphics and with numerical values in
percentage or absolute value, (see § 0 as below shown:

12/09/2006 – 16:55:10

TOTAL HARM VOLTAGE – Page 1/5

h 01

Fs 100%
V1h

100.0 %

V2h

100.0 %

V3h

100.0 %

Vnh

0.0 %

Thd

6.5 %

Thd

5.9 %

Thd

4.3 %

Thd

0.0 %

PAGE

HARM I

TABLE

HARM PAG

12/09/2006 – 16:55:10

TOTAL HARM VOLTAGE – Page 1/5

h 01

Fs 400.0V
V1h

228.6V

V2h

225.1V

V3h

230.7V

Vnh

0.0V

Thd

6.5 %

Thd

5.9 %

Thd

4.3 %

Thd

0.0 %

PAGE

HARM I

TABLE

HARM PAG

Fig. 53: Harmonic analysis of voltages in percentage/absolute values for 4-wire systems

12/09/2006 – 16:55:10

TOTAL HARM VOLTAGE – Page 1/4

h 01

Fs 100%
V12h

100.0 %

V23h

100.0 %

V31h

100.0 %

Thd

6.5 %

Thd

5.9 %

Thd

6.5 %

PAGE

HARM I

TABLE

HARM PAG

12/09/2006 – 16:55:10

TOTAL HARM VOLTAGE – Page 1/4

h 01

Fs 400.0V
V12h

328.4V

V23h

389.6V

V31h

386.6V

Thd

6.5 %

Thd

5.9 %

Thd

4.3 %

PAGE

HARM I

TABLE

HARM PAG

Fig. 54: Harmonic analysis of voltages in percentage/absolute values for 3-wire and ARON

12/09/2006 – 16:55:10

PHASE1 HARM VOLTAGE – Pag 1/2

h 01

Fs 100%
V1h

100.0 %

Thd

6.5 %

PAGE

HARM I

TABLE

HARM PAG

12/09/2006 – 16:55:10

PHASE1 HARM VOLTAGE – Pag 1/2

h 01

Fs 400.0V
V1h

228.6V

Thd

6.5 %

PAGE

HARM I

TABLE

HARM PAG

Fig. 55: Harmonic analysis of voltages in percentage/absolute values for Single phase


In all case the full scale of graphs is automatically set from meter depending on measured
values.