Vega78, Fig. 99: voltage anomalies selection screen – HT instruments VEGA78 User Manual

Page 53

background image

VEGA78

EN - 51

Press SAVE or ENTER keys (or the smart icon

) to save each selection and confirm by

“Ok”. The main screen of Fig. 85 is shown by meter at the end of the operation.
Press ESC key (or the smart icon ) to quit without saving and back to previous screens.
The selection parameters of harmonic analysis require the preliminary selection of
voltages or currents inside General Parameter sub-level. The below error screen are
shown in these cases:

12/09/2006 – 16:55:10

RECORDING SETTINGS

- Harmonics

+

Voltages

+

Currents

Voltage Anom.: 230V 6%-10%

Unbalance

Error: no selected current

CLP/EXP

TYPICAL

SEL

MODIFY

Fig. 97: No selected currents error screen

To solve the error (see Fig. 97) situation of above screen select the “Current” inside
“General Parameter” sub-level (see § 5.4.6.1).

12/09/2006 – 16:55:10

RECORDING SETTINGS

- Harmonics

+

Voltages

+

Currents

Voltage Anom.: 230V 6%-10%

Unbalance

Error: no selected voltage

CLP/EXP

PREDEF.

SEL

MODIFY

Fig. 98: No selected voltages error screen

To solve the error (see Fig. 98) situation of above screen select the “Voltages” inside
“General Parameter” sub-level (see § 5.4.6.1).

5.4.7. Voltage Anomalies
This option permits to set the control parameters relative to voltage anomalies (sags,
swell, supply breaks – see § 10.1) recording which is completely independent from
periodic analysis (regulated by integration period). The below screen is shown by meter:

12/09/2006 – 16:55:10

RECORDING SETTINGS

+ General Parameters

+

Harmonics

Voltage Anom.: 230V 6%-10%

Unbalance

132/251 Parameters – Autonomy: 15d 12h

CLP/EXP

PREDEF.

SEL

MODIFY

Fig. 99: Voltage Anomalies selection screen